OPERATION MANUAL
NAV-402AP-3
1-2-1
Page 2
Sep 1/04
1.5 AC POWER
REQUIREMENTS
The Test Set operates over a voltage range of
100 to 120 VAC at 60 Hz or 220 to
240 VAC at 50 Hz according to VOLTAGE
SELECT Switch setting.
The specified fuse ratings are listed in
1-2-1, Table 1.
CAUTION: FOR CONTINUOUS
PROTECTION AGAINST FIRE,
REPLACE ONLY WITH FUSES
OF THE SPECIFIED VOLTAGE
AND CURRENT RATINGS.
INPUT VOLTAGE F1 AND F2 FUSES
100 to 120 VAC 1.0 A, 250 V
Fast Blo (Type F)
(Aeroflex 5106-4501-000)
(1.2in AGC Glass)
220 to 240 VAC 0.5 A, 250 V
Fast Blo (Type F)
(Aeroflex 5106-0000-016)
(1.2in AGC Glass)
Specified Fuse Ratings
Table 1
1.6 FCC REQUIREMENTS
The Test Set is prohibited from use in VAR
(variable) frequency modes during ramp
use. Any test signals radiated from the
antenna must be fixed frequencies. The
Test Set is prohibited from externally
modulating any band during ramp use.
This includes microphones, audio
generators, etc.
1.7 BATTERY RECHARGING
Refer to 1-2-1, Figure 1.
STEP PROCEDURE
1. Verify FUSES are correct for normal
operating voltage. Refer to para 1-2-1.5.
2. Verify VOLTAGE SELECT Switch is set to
the setting (115 or 230 VAC) that
matches the input ac power.
3. Connect ac power cable between AC
Power Connector and normal operating
voltage power source, according to Test
Set configuration. Refer to para 1-2-1.5.
NOTE: The battery charger operates
whenever ac power is applied to
the Test Set.
4. Allow several hours for battery charge or
until 0 to 30% scale displays battery
voltage is 15 V or greater.
NOTE: Overnight charging is
recommended.