OPERATION MANUAL
NAV-402AP-3
1-2-2
Page 6
Sep 1/04
ITEM DESCRIPTION
17. AVG/PEAK LOC XMTR ON Switch
Selects power meter function displayed
on MONITOR Meter when MONITOR
METER FUNCTION Switch is set to 10W
or 100W:
O PEAK sets the MONITOR Meter to
display peak power amplitude of a
COM Transmitter with modulation
applied. PEAK also provides
Localizer Transmitter output
simultaneously with the Glide Slope
signal when the FREQUENCY MODE
Switch is set to G/S FIXED (for
checking Autopilots or Course
Computers).
O AVG sets the MONITOR Meter for
displaying the average RF power of a
COM Transmitter and disregards
power changes due to proper
modulation.
18. 90/150 Hz VAR PHASE (φ) Switch
Varies the phase (φ) of the 150 Hz tone
relative to the 90 Hz tone. Normally, the
90 and 150 Hz tones, used in Glide
Slope and Localizer simulations, are
locked in phase together. Adjusting the
VOR BRG SELECT Switches while
pressing the 90/150 Hz VAR PHASE (φ)
Switch, changes phase relationship. If
the VOR BRG SELECT Switches are set
to 000.0, pressing the 90/150 Hz VAR
PHASE (φ) Switch produces no change in
phase. Each 1° step selected on the
VOR BRG SELECT Switches shifts the
150 Hz tone 5° relative to the 90 Hz
tone. The 90 Hz tone always remains
constant in phase relative to the 000.0
set on the VOR BRG SELECT Switches.
Setting the VOR BRG SELECT Switches
to 024.0 (24° VOR bearing) with the
90/150 Hz VAR PHASE (φ) Switch
pressed puts the 90 and 150 Hz tones in
original phase relationship. Setting the
VOR BRG SELECT Switches to 012.0
(12° VOR bearing) with the 90/150 Hz
VAR PHASE (φ) Switch pressed, places
the amplitude summation of the 90 and
150 Hz tones at maximum magnitude.
ITEM DESCRIPTION
19. G/S CENTER Adjust
Varies the amplitude of the 90 Hz tone
and controls exact course centering for
Glide Slope modulation.
20. G/S VAR DDM Control
Continuously varies G/S pointer
deflection from counterclockwise (ccw)
0.8 DDM (down) through 0 DDM to
clockwise (cw) 0.8 DDM (up). Setting the
G/S DDM Control to p activates the G/S
VAR DDM Control.
21. TONE SELECT Switch
Selects modulation input to the internal
Signal Generator or input to the internal
digital Counter:
O CTR IN connects the MOD IN/EXT
FREQ CTR IN Connector input to the
Counter, active with COUNTER
MODE Switch set to HZ, KHZ, KHZ or
MHZ.
O MOD IN applies the MOD IN/EXT
FREQ CTR IN Connector modulation
to the Signal Generator.
O 1020 applies 1020 Hz modulation to
the frequency band (except G/S) set
by FREQUENCY MODE Switch.
O 400 applies 400 Hz modulation to the
frequency band (except G/S) set by
FREQUENCY MODE Switch.
O 1300 applies 1300 Hz modulation to
the frequency band (except G/S) set
by FREQUENCY MODE Switch.
O 3000 applies 3000 Hz modulation to
the frequency band (except G/S) set
by FREQUENCY MODE Switch.
22. MOD IN/EXT FREQ CTR IN Connector
Provides an input for external modulation
to the internal Signal Generator or a 1 to
300 MHz signal to the internal digital
Counter as selected by the TONE
SELECT Switch.